How to calibrate the Mettler weighing terminal IND320?
Standard Calibration
Standard Calibration uses two points,zero and span point.Steps are as follows:
[System Setup]→[Scale Setup]→[Set Capacity].Set the scale capacity in this item.
[System Setup]→[Scale Setup]→[Set Capacity].Set the increment in this item.
[System Setup]→[Scale Setup]→[Select Units].Select the scale unit in this item.
1.Adjust Zero
Step 1:Access to the zero capture menu, [System Setup]→[Scale Setup]→[Adjust Zero].
Step 2:Press Enter key, the IND320 displays “Adjust Zero?”.Press Select key to choose “√”,and then press Enter key to perform zero adjust.
Step 3:The IND320 displays “Empty?” to inform the users to empty the scale.
Step 4:Press Enter key,the IND320 performs zero capture.The IND320 displays “Adjust Zero 20%”.The guage ‘20%’, ‘40%’...... ‘100’ indicates the status of the adjustment.
2.Adjust Span
Step 1:Access the span capture menu, [System Setup]→[Scale Setup]→[Adjust Span]
Step 2: Press Enter key, the IND320 displays “Adjust Span”.Press Select key to choose “√”,and then press Enter key to perform span adjust.
Step 3:Press Enter key,when the IND320 displays “Load Weigh”.Input the load weight in “Input Weigh 000000”,and press Enter key to perform span capture.
1.Adjust Zero
Step 1:Access to the zero capture menu, [System Setup]→[Scale Setup]→[Adjust Zero].
Step 2:Press Enter key, the IND320 displays “Adjust Zero?”.Press Select key to choose “√”,and then press Enter key to perform zero adjust.
Step 3:The IND320 displays “Empty?” to inform the users to empty the scale.
Step 4:Press Enter key,the IND320 performs zero capture.The IND320 displays “Adjust Zero 20%”.The guage ‘20%’, ‘40%’...... ‘100’ indicates the status of the adjustment.
2.Adjust Span
Step 1:Access the span capture menu, [System Setup]→[Scale Setup]→[Adjust Span]
Step 2: Press Enter key, the IND320 displays “Adjust Span”.Press Select key to choose “√”,and then press Enter key to perform span adjust.
Step 3:Press Enter key,when the IND320 displays “Load Weigh”.Input the load weight in “Input Weigh 000000”,and press Enter key to perform span capture.
1.Adjust Zero
Step 1:Access to the zero capture menu, [System Setup]→[Scale Setup]→[Adjust Zero].
Step 2:Press Enter key, the IND320 displays “Adjust Zero?”.Press Select key to choose “√”,and then press Enter key to perform zero adjust.
Step 3:The IND320 displays “Empty?” to inform the users to empty the scale.
Step 4:Press Enter key,the IND320 performs zero capture.The IND320 displays “Adjust Zero 20%”.The guage ‘20%’, ‘40%’...... ‘100’ indicates the status of the adjustment.
2.Adjust Span
Step 1:Access the span capture menu, [System Setup]→[Scale Setup]→[Adjust Span]
Step 2: Press Enter key, the IND320 displays “Adjust Span”.Press Select key to choose “√”,and then press Enter key to perform span adjust.
Step 3:Press Enter key,when the IND320 displays “Load Weigh”.Input the load weight in “Input Weigh 000000”,and press Enter key to perform span capture.